
Warning:
To avoid spreading contamination with biologically hazardous, toxic, and
corrosive materials, dispose of all waste materials according to local environmental
regulations.
5. Remove and dispose of the old filter.
6. Place the new filter flat on the inside part of the grill, with its edges beneath the metal lip.
7. Close the air filter grill.
8. Connect the probe cable to the high-voltage connector.
9. Close the source enclosure.
10. Lower the instrument's visor.
4.23 Replacing the UniSpray probe assembly
Replace the probe assembly on the UniSpray source if it becomes irreversibly blocked, or if it
becomes contaminated or damaged.
4.23.1 Removing the UniSpray probe assembly
Required materials
• Chemical-resistant, powder-free gloves
Warning:
To avoid personal contamination with biologically hazardous, toxic, and corrosive
materials, wear chemical-resistant, powder-free gloves when performing this procedure.
To remove the UniSpray probe assembly:
1. To prepare the instrument for working on the source, stop solvent flow, ensure that the
instrument is in Standby mode, and stop desolvation gas flow.
Warning:
To avoid burn injuries, exercise care when handling the components
of the source enclosure heated to high temperatures. Wait until the hot
components have sufficiently cooled before you handle them.
2. Unscrew the probe's PEEK fitting to disconnect the probe assembly's input from the
fluidics.
January 10, 2022, 715006200 Ver. 01 (previously released as Rev. A)
Page 135