Removal and refitting of the source enclosure
4-9
6. Using both hands, grasp the source enclosure, and lift it vertically off the
two supporting studs on the source adaptor housing.
Fitting the source enclosure to the instrument
Required material
Chemical-resistant, powder-free gloves
To fit the source enclosure
1. Using both hands, fit the source enclosure to the two supporting studs
on the source adaptor housing.
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
TP03164
Supporting stud
Source enclosure