Preparing the IntelliStart Fluidics system
2-7
To install the low-volume vials
1. If a standard reservoir bottle is fitted, remove it.
2. Screw the low-volume adaptors into the manifold, and finger tighten
them.
3. Screw the low-volume vials into the adaptors.
Purging the infusion pump
Whenever you replace a solution bottle, purge the infusion pump with the
solution that you are going to use next. See the mass spectrometer’s online
Help for details.
Tip:
Depending on the solutions used, the instrument’s solvent delivery
system can require more than one purge cycle to minimize carryover.
Warning:
The reservoir bottles can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Low-volume vial
Low-volume adaptor