MS/MS operating modes
1-11
MS/MS operating modes
Note:
In SIR and MRM modes, neither quadrupole is scanned, so no spectrum
(intensity versus mass) is produced.
The following table shows the MS/MS operating modes.
MS/MS operating modes
Operating mode
MS1
Collision cell
MS2
Product
(daughter) ion
spectrum
Static (at
precursor mass)
Pass all masses
Scanning
Precursor
(parent) ion
spectrum
Scanning
Static (at product
mass)
MRM
Static (at
precursor mass)
Static (at product
mass)
Constant neutral
loss spectrum
Scanning
(synchronized
with MS2)
Scanning
(synchronized
with MS1)
ScanWave
daughter scan
Static (at
precursor mass)
ScanWave
enabled
Scanning
(synchronized
with collision
cell)