Optional glass capillary sprayer
4-11
3. Retract the sprayer platform adjuster assembly from the source (see
).
4. Remove the sprayer’s safety cover.
5. Place the sprayer (with gas line fitted) on the platform, and secure it
with the thumbscrew.
6. Refit the safety cover.
Fitting and loading the glass capillary
Required materials
•
Chemical-resistant, powder-free gloves
•
Fused silica syringe needle or a GELoader
®
tip
To fit and load the glass capillary:
1. Unscrew the union at the front of the sprayer, and remove the sprayer’s
front section.
f
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
PTFE backpressure gas tubing
Blue conductive elastomer
Ferrule
Glass capillary
Knurled nut
Union