5-102 Maintenance Procedures
Replacing the mass spectrometer’s air filters
Replacing the air filter inside the front door
Required materials
•
Needle-nose pliers
•
New filter
To replace the air filter inside the front door:
1. Open the access door to the fluidics pump (see the figure on
).
2. Unscrew the captive thumbscrew on the filter cover.
Thumbscrew
Filter cover