5-84
Maintenance Procedures
8. Unscrew the finger-tight nut on the rear of the reference sprayer
assembly.
9. Remove the liner tubing and capillary from the reference sprayer
assembly.
10. Remove the capillary from the liner tubing.
11. Dispose of the capillary in accordance with local environmental
regulations.
Installing the new capillary
Required materials
•
Chemical-resistant, powder-free gloves
•
4-mm wrench
•
Spares kit for reference probe assembly
•
Extender tool for the SealTight nut (included in the spares kit)
To install the new capillary:
1. Fit the capillary into the liner tubing.
2. Fit the liner tubing and capillary to the reference sprayer assembly.
3. Fit the reference probe tip and, using the 4-mm wrench, screw it in until
the end of the capillary protrudes 0.5 to 1.0 mm from the probe tip.
4. Tighten the “finger-tight” nut on the rear of the reference sprayer
assembly.
5. Insert the reference sprayer assembly into its support assembly,
with
the sprayer tip pointing upwards.
Warning:
The capillary can be contaminated with
biohazardous and/or toxic materials. Dispose of it according
to local environmental regulations.
Caution:
To prevent pressure from ejecting the capillary from the
sprayer, firmly tighten the finger-tight nut.
Summary of Contents for Xevo G2 QTof
Page 10: ...x...
Page 18: ...xviii Table of Contents...
Page 34: ...1 16 Waters Xevo G2 QTof Overview...
Page 44: ...2 10 Preparing the Mass Spectrometer for Operation...
Page 227: ...Connecting the liquid waste line B 17 Positioning of drain tube TP01807 Correct Incorrect...
Page 238: ...B 28 External Connections...
Page 242: ...C 4 Materials of Construction and Compatible Solvents...
Page 248: ...Index 6...