4-8
Operating the Waters TQ Detector
2.
In the ACQUITY UPLC Console system tree, click TQ Detector.
3.
Ensure that the Ion Mode is ES+.
4.
In the ACQUITY UPLC Console system tree, click IntelliStart.
5.
Clear the Instrument Setup check box.
6.
If you require system pre-checking, click Pre-checks.
Rationale:
If Pre-checks is selected, when IntelliStart starts the sample
tune, it determines whether the existing tune is still valid. If so, no
sample tune is performed.
See also:
TQ Detector online Help topic “IntelliStart flow diagram”.
7.
Click Start
.
8.
In the IntelliStart Setup Parameters dialog box’s Sample Tune tab,
specify the sample tune parameters as described in
sample tune parameters:” on page 4-5
9.
Click Start
.
Result:
A message appears reminding you to ensure that the tune
solutions, tune parameters, and LC flow are set correctly.
10. Click OK.
Result:
Sample tune starts. The ACQUITY UPLC Console displays the
progress of the tune. When the sample tune is complete, the results are
written to the Sulfadimethoxine.ipr file.