3-10
ESI and ESCi Modes of Operation
4.
Use the needle-nose pliers to remove the corona pin from its mounting
contact.
Tip:
Store the corona pin in a safe location.
See also:
Figure titled “Corona pin:” on page 3-7
5.
Use the needle-nose pliers to fit the blanking plug to the corona pin
mounting contact.
See also:
Figure titled “Corona pin mounting contact:” on page 3-6
6.
Close the source enclosure door and fasten the handle.
7.
Close the instrument’s access door.
Warning:
The corona pin tip is sharp. To avoid puncture wounds,
handle the corona pin with care.
Caution:
To avoid damaging to the corona pin’s tip and bending
the pin, use the needle nose pliers to grip the corona pin at the
end that fits into the mounting contact.