1-18
Waters TQ Detector Overview
6.
Click OK.
7.
Click Reservoir A or B.
8.
In the Select Reservoir dialog box, click A or B.
9.
Click OK.
Tip:
If the selected reservoir is different from the current reservoir, the
system purges as the reservoir changes.
10. Click the current diverter valve position label.
11. In the Select a Flow State dialog box, select the required flow state.
12. Click OK.
13. Click
to
start.
Tip:
A status indicates the amount of fluid in the syringe and the
amount of time remaining before the fluid empties. When the syringe is
empty the system becomes idle.
14. Click
to refill the syringe or
to purge the system.
15. Click
to stop the current action.
Operating the IntelliStart fluidics system from the Tune window
To operate the IntelliStart fluidics system from the Tune window:
1.
In the MassLynx window, click MS Tune.
2.
In the TQ Detector Tune window, click the Fluidics page.
3.
Set the Flow Control parameters according to the instructions in the
MassLynx online Help.
Programming the MS method to operate the IntelliStart fluidics
system
In the MS method, you can program the operation of the system’s
multi-position valve to infuse sample during a run. The valve can also divert
LC flow to waste as a timed event.