IntelliStart Fluidics system
1-9
Dual-mode ionization source
Atmospheric pressure photoionization (APPI) uses photons generated by a
discharge UV lamp (~10.2 eV) to produce sample ions from vaporized LC
eluent. Direct photoionization of the sample molecule occurs when the photon
energy exceeds the ionization potential of the sample molecule.
The optional dual-mode (APPI/APCI) ionization source incorporates an APPI
source enclosure used in conjunction with a standard APCI probe. You can
operate the source in APPI, APCI, or dual-mode, which switches rapidly
between ionization modes, facilitating high-throughput analyses.
See also:
The
Waters SYNAPT G2 Dual-mode Ionization Source Operator’s
Guide
.
Matrix-assisted laser desorption ionization
The matrix-assisted laser desorption ionization (MALDI) interface enables
rapid, tool-free switching between API and MALDI modes. A motorized stage
moves the MALDI source into position.
See also:
The
Waters MALDI SYNAPT G2 MS System Operator’s Guide
.
IntelliStart Fluidics system
Overview
The IntelliStart Fluidics system is built into the instrument; it controls how
sample is delivered to the source.
For standard flow applications, the system delivers sample directly to the
mass spectrometer source in one of three ways:
•
From the LC column.
•
From three integral vials.
Tip:
The vials can also deliver sample through direct or combined
infusion to enable optimization at analytical flow rates.
•
From a wash reservoir that contains solvent for flushing the
instrument’s solvent delivery system.
Summary of Contents for SYNAPT G2
Page 18: ...xviii Table of Contents...
Page 46: ...2 8 Starting Up and Shutting Down the Mass Spectrometer...
Page 66: ...3 20 Configuring the LockSpray Source...
Page 228: ...B 24 External Connections 7 Click Next Result The software installs 8 Click Finish...
Page 232: ...C 4 Materials of construction and compliant solvents...