Changing the UV lamp bulb
4-5
APPI lamp drive assembly
To change the UV lamp bulb
1. Remove the APPI lamp drive assembly from the source enclosure.
“To remove the APCI probe and APPI lamp drive assembly
from the instrument” on page 2-8
through
.
2. Use the cross-head screwdriver to remove the repeller electrode
retaining screw.
3. Remove the repeller electrode from the PEEK™ insulator.
Warning:
To avoid self contamination with toxic and
biohazardous materials, always wear chemical-resistant,
powder-free gloves at all times while handling the
components. The probe, source, and APPI lamp drive
assembly may be contaminated with biologically hazardous
materials.
Warning:
To avoid electric shock, ensure that the instrument is
in standby before commencing this procedure.
PEEK™ insulator
Repeller electrode
retaining screw
Repeller electrode
Insulator retaining
screw (2-off)
Source enclosure side plate