Operating in Dual APPI/APCI mode
3-15
8. Use the needle-nose pliers to fit the dual APPI/APCI corona pin into the
mounting contact. Ensure that the pin tip aligns with the sample cone
tip.
Combined APPI/APCI corona pin alignment
Note:
For clarity, the lamp drive assembly is not shown.
9. Close the source enclosure door, and fasten the securing clips.
10. Close the instrument’s front access door.
To confirm that the UV lamp is operating correctly
1. On the MassLynx Tune window, click Ion Mode > APcI+ or Ion Mode >
APPI+.
2. Click Options > Multiple Ion Modes >Dual APPI/APcI.
APCI probe
Sample cone tip
Combined
APPI/APCI
corona pin