
3-14
Operating in APPI, APCI, and Dual APPI/APCI Modes
4. Unfasten the source enclosure door’s securing clips, and open the door.
5. Use methanol or isopropyl alcohol, applied to the lint-free cloth, to
carefully clean the lamp window.
6. If no corona pin is fitted, use the needle-nose pliers to remove the
blanking plug from the corona pin mounting contact.
Tip:
Store the blanking plug in a safe location.
7. If the APCI corona pin is fitted, use the needle-nose pliers to remove the
pin.
Tip:
Store the APCI corona pin in a safe location.
Warning:
To avoid self-contamination with toxic and
biohazardous materials, always wear chemical-resistant,
powder-free gloves at all times while handling the
components.
Warning:
To avoid electric shock, ensure that the instrument is
in standby while working with the instrument’s front access door
open.
Warning:
To avoid burn injuries, take great care while working
with the instrument’s front access door open. The probe and
source may be hot.
Caution:
To avoid damage, do not apply any force to the source
enclosure door while the door is open.
Summary of Contents for Quattro Premier XE
Page 14: ...4 Table of Contents...
Page 64: ...4 16 Maintenance Procedures...
Page 73: ...Warnings that apply to all Waters instruments A 9...
Page 82: ...B 4 Materials of Construction and Compliant Solvents...
Page 86: ...Index 4...