Operating in APCI mode
3-11
c.
Click
to stop the nitrogen flow.
4. Unfasten the source enclosure door’s securing clips and open the door.
5. If no corona pin is fitted, use the needle-nose pliers to remove the
blanking plug from the corona pin mounting contact.
Tip:
Store the blanking plug in a safe location.
6. If the dual APPI/APCI corona pin is fitted, use the needle-nose pliers to
remove the pin.
Tip:
Store the dual APPI/APCI corona pin in a safe location.
Warning:
To avoid self-contamination with toxic and
biohazardous materials, always wear chemical-resistant,
powder-free gloves while handling the components.The
probe and source may be contaminated with biologically
hazardous materials.
Warning:
To avoid electric shock, ensure that the instrument is
in standby while working with the instrument’s front access door
open.
Warning:
To avoid burn injuries, take great care while working
with the instrument’s front access door open. The probe and
source may be hot.
Caution:
To avoid damage, do not apply any force to the source
enclosure door while the door is open.