2-38
Installing the System
System column rack (Delta-Prep system)
Before you connect to the column rack, consult the table below for material
types and sizes.
Column rack connections
Column rack
connection
Type of
tubing
Tubing ID
Connect to
Inlet from
pump
Stainless
0.040
Transducer outlet on pump
Column outlet,
small scale
Stainless
0.020
Detector inlet
Column outlet,
large scale
Stainless
0.040
Detector inlet
Recycle to
pump
PTFE
0.040
Recycle port on manual sample
inlet valve
Inlet from
detector
Stainless
0.040
Detector outlet
TP014
7
1
Waters
D
e
l
taPrep
Inlet (from pump)
Cabinet latch
Inlet (from detector)
Small-scale column outlet
Large-scale column outlet
Recycle to pump
Fraction outlets selector
Fraction outlets 1-8
Purge outlet
Vials
Fractions/waste/recycle valve
Purge/ system valve
Waste
Small/large-scale
switching valve
Large-scale
injector
Small-scale
injector