Item
Specification
Temperature stability Tested to ±0.3 °C
Test conditions:
• Door closed
• No column installed
• No flow
• Measurement taken with traceable, external temperature
measurement device
• Measurement taken for 1 hour after thermal equilibration at set point
• Measurement taken at column compartment sensor location
• Tested at 35 °C, 60 °C, and 90 °C
Ambient temperature
stability
Within 2.0 °C/60 minute maximum
Pre-heater
temperature (not
user-settable)
Defined by set point of the column zone
Solvent conditioning Active pre-heating
Leak control
Compartment drip tray with leak sensor installed under compartment.
Single exit drain manages leaks to waste.
Column tracking
eCord technology column information management tracks and archives
column usage history
a. Operating with active pre-heating via the APH assembly
C.2.7 Sample organizer performance specifications
Item
Specification
Sample plate
compatibility
Consult the sample manager’s overview and maintenance guide or
system consumables catalog for a list of approved plates, vials,
sealing caps, and sample covers for use with your sample manager
and sample organizer.
Sample plate capacity
Maximum of 19 plates, as high as 15.5 mm
Maximum of 9 plates, as high as 40.0 mm
Maximum of 6 plates, as high as 53.0 mm
Minimum sample plate
height
13 mm
August 8, 2016, 715005049 Rev. C
Page 115