Probe entrance
Cable screws
Source enclosure thumbscrew
Source enclosure thumbscrew
6.
Remove the source enclosure by pulling it away from the instrument using both hands.
Warning:
To avoid puncture injuries from the sharp probe tip, use care when
inserting and removing the probe from the source enclosure.
7.
Remove the probe from the top of the source enclosure using the probe fitting (see the
figure "Inserting the probe" in
).
3.7.2 Fitting the source enclosure to the instrument
Required materials
• Chemical-resistant, powder-free gloves
Warning:
To avoid personal contamination with biohazardous or toxic compounds, wear clean,
chemical-resistant, powder-free gloves when performing this procedure.
To fit the source enclosure:
1.
Use both hands to slide the source enclosure onto the instrument's supporting rods.
2.
To secure the enclosure against the instrument, tighten the 2 thumbscrews on the front of
the source enclosure.
Notice:
To avoid damage to the electrical connector's screws, do not
overtighten them.
3.
Connect the electrical cable to the socket on the right-hand side of the instrument's front
panel and tighten the screws.
Notice:
To avoid damaging the fragile probe, use care when inserting it into the
source enclosure.
4.
Fit the probe to the source enclosure (see
).
June 15, 2017, 715003956 Rev. F
Page 30