26
I/O signal connector
The detector’s rear panel includes a removable connector that holds the screw
terminals for I/O signals. This connector is keyed so that it can receive a
signal cable inserted only one way.
PDA I/O signal connector:
Connecting to the electricity source
The ACQUITY UPC
2
PDA detector requires a separate, grounded electricity
source. The ground connection in the electrical outlet must be common and
connected near the system.
ACQUITY UPC
2
PDA detector analog/event connections:
Signal connections
Description
Inject Start
Start injection
Event Out
Output switch to trigger external devices
Analog Out
Analog chart output
Warning:
To avoid electrical shock, observe these precautions:
• Use power cord SVT-type in the United States and HAR-type (or
better) in Europe. For other countries, contact your local Waters
distributor.
• Power-off and unplug the detector before performing any
maintenance on the instrument.
• Connect all components of the ACQUITY UPC
2
system to a common
ground.
Inject Start
Event Out
Analog Out
1
2
3
4
5
6
+
–
+
–
+
–
Summary of Contents for ACQUITY UPC2