9. Refit the PEEK cone clamp and the gas cone assembly to the ion block.
10. Fit the source enclosure to the instrument (see
Fitting the source enclosure to the
3.11 Replacing the source enclosure
Required tools and materials
Warning:
To avoid personal contamination with biologically hazardous or toxic compounds,
wear clean, chemical-resistant, powder-free gloves when performing this procedure.
Notice:
To avoid damaging the fragile probe, use care when removing it from the
source enclosure.
Warning:
To avoid puncture injuries from the sharp ESI capillary, use care when
inserting and removing the probe from the source enclosure.
Warning:
To avoid burn injuries, exercise care when handling the components of the
source enclosure heated to high temperatures. Wait until the hot components have
sufficiently cooled before you handle them.
Important:
To avoid injury, do not remove the ion block, restrictor, or source ion guide when the
instrument is in Power Save mode. To remove these components you must turn off the
instrument and wait five minutes for it to vent. Power-down the instrument by pressing and
holding the power button for approximately five seconds until the instrument powers-down.
Required materials
• Chemical-resistant, powder-free gloves
• Replacement source enclosure
To replace the source enclosure:
1. In the software, click
Power Save
to put the instrument into Power Save mode.
2. Open the instrument door.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)
Page 50