2 Preparing for operation
To confirm that the instrument is ready for operation, verify the following conditions:
• The power supply, the vacuum pump, and the nitrogen gas are connected to the instrument.
for details on all the instrument's external connections.
• The following fluidics connections are made:
• The LC outlet flow (from the column or optical detector) is connected to the divert valve.
• The ESI capillary is connected between the divert valve and ESI source inlet.
• The wash, lock mass, and calibration bottles all contain a sufficient volume of solution and
are mounted to the instrument in the correct reservoir slots. See
.
• The instrument door is closed.
If you moved the instrument and you need to reconnect it, see
2.1 Software and data system
UNIFI software, which controls the ACQUITY RDa Detector and LC system, enables the
following major operations:
• Configuring the system
• Using the setup feature to automatically tune and mass calibrate the detector
• Monitoring the analysis
• Acquiring data:
• Full Scan
• Full Scan with Fragmentation
• Processing data
• Reviewing data
• Printing reports
2.1.1 Calibrating and tuning the instrument
To calibrate and tune the instrument, use the Instrument Setup function on the
Acquity RDa
tab
to define the mass range and polarity settings, and then click
Start
.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)