Table D–1: I/O signal connections (continued)
Callout
Screenprint
Software function
Description
ACQUITY UPLC I-
Class/Premier BSM.
For the ACQUITY RDa
Detector, use Pins 1
and 2; for the
ACQUITY UPLC I-
Class/Premier BSM,
use Pins 3 and 4.
Maximum 30 V, 0.5 A
Event Out 2
Used to send time-
based contact closure
signals to external
devices. Maximum 30
V, 0.5 A
Switch -
Injection Start +
Event In 1
Signals the start of an
injection. >5 V or <1.5
V
Injection Start -
Event In Ground
Event In +
Event In 2
Allows an external
device to start data
acquisition. >5 V or
<1.5 V
Event In -
Event In Ground
Analog -
Analog Ground
0 to 10 V, not used
Analog Output
Note:
Event Out 1, Event Out 2, Event In 1, and Event In 2 are configurable in the software.
D.8.1 Signal connections
Requirement:
To meet the regulatory requirements of immunity from external electrical
disturbances, install connection covers over the signal connectors.
To make signal connections:
1. For input and output signal connection details, see
Input/output signal connectors (Page
2. To make the signal connections, attach the positive and negative leads of the signal cable
to the connector.
August 20, 2021, 715005003 Ver. 04 (Previously released as Revision C)