3 Maintenance Procedures
84
September 26, 2013, 715004208 Rev. A
To replace the air filter:
1. Remove the old air filter from the air filter frame, and discard it.
2. Align the new air filter within the air filter frame.
Replacing the optional leak sensor
Required materials
•
Gloves: clean, powder-free, chemical-resistant
•
Leak sensor
Warning:
To prevent injury, always observe Good Laboratory Practices
when you handle solvents, change tubing, or operate the ISM. Consult
the Material Safety Data Sheets regarding the solvents you use.
Warning:
To avoid personal contamination with biohazards,
compounds that are toxic or corrosive, or the residues that
make up such compounds, wear clean, chemical-resistant,
powder-free gloves when handling the leak sensor.
Frame
Air filter