Resolving leak sensor errors
September 26, 2013, 715004208 Rev. A
35
8. Plug the leak sensor connector into the receptacle on the front of the
instrument.
9. Power-on the ISM.
10. In the Instrument console, select Isocratic Solvent Manager from the
system tree.
11. In the ISM information window, click Control > Reset ISM, to reset the
ISM.
12. In the console, enable the leak sensor, to activate the leak sensor’s leak
detection capability.
Resolving leak sensor errors
After approximately 1.5 mL of liquid accumulates in the leak sensor reservoir,
an alarm sounds, indicating that the leak sensor detected a leak.
Warning:
To prevent injury, always observe Good Laboratory Practices
when you handle solvents, change tubing, or operate the ISM. Consult
the Material Safety Data Sheets regarding the solvents you use.
Warning:
To avoid personal contamination with biohazards,
compounds that are toxic or corrosive, or the residues of such
compounds, wear clean, chemical-resistant, powder-free
gloves when reinstalling fittings.
Caution:
To avoid scratching or otherwise damaging the leak sensor,
• do not allow buffered solvents to accumulate and dry on it;
• do not submerge it in a cleaning bath.
Leak sensor
connector receptacle