Input/output specifications
September 26, 2013, 715004208 Rev. A
155
Input/output specifications
The following table lists the input and output specifications for the ISM.
Performance specifications
ISM
The following table lists the performance specifications for the ISM.
d.
Moisture Protection
– Normal (IPX0) – IPX0 means that no Ingress Protection against any
type of dripping or sprayed water exists. The “X” is a placeholder that identifies protection
against dust, if applicable.
Input and output specifications:
Attribute
Specification
Gradient start
Maximum input voltage: ±30 VDC
Logic High:
≥
3.0 VDC
Logic Low:
≤
1.9 VDC
Minimum pulse width: 100 msec
Stop flow
Maximum input voltage: ±30 VDC
Logic High:
≥
3.0 VDC
Logic Low:
≤
1.9 VDC
Minimum pulse width: 100 msec
ISM performance specifications:
Item
Specification
Priming
Automated priming at flow rates of
4 mL/min.
Check valves
Primary and accumulator check valves
are passive
Vent valve
Automated vent valve for priming and
automated leak testing
Number of solvents
One