December 5, 2017, 715004753 Rev. B
Page 49
Figure 3–2:
Installing the analytical flow cell assembly:
10. Continue to insert the flow cell until the three thumbscrews align with their holes in the
bulkhead.
Note:
To prevent the flow cell from binding and to ensure that it is properly seated in the
bulkhead, alternate between tightening the captive screws and pushing the flow cell
forward.
11. Hand tighten the thumbscrews to the extent possible, and then tighten slightly more using
the 1/4-inch flat-blade screwdriver.
12. Connect the column-outlet tubing to the flow cell inlet
.
13. Connect the waste line to the flow cell outlet.
14. Close the door.
15. Before you power-on the detector, prime the system to fill the flow cell with solvent and
remove any air.
3.5
Replacing the lamp
Change the lamp when it repeatedly fails to ignite or when the 2998 PDA detector fails to calibrate.
The detector’s source lamp is warranted to light and pass startup diagnostic tests for 2000 hours
or 1 year from the date of purchase, whichever event occurs first.
Alignment pin
Flow cell handle