June 11, 2015, 715004752 Rev. A
Page 73
To program a new timed event:
1. Press the METHOD (Shift A/B) key on the detector keypad.
Figure 3–17: Method choice list
2. Press 1, Timed events.
3. Enter the time for the event.
Figure 3–18: Timed events screen
4
Chart-mark
(10% of full
scale)
Does not
apply
Does not apply
Yes
5
Polarity
1. +
2. –
+
Yes
6
Auto-zero
Does not
apply
Does not apply
Yes
7
Lamp
1. Off
2. On
Off
No
8
Switch 1
1. High
2. Low
3. Pulse
4. Rect wave
High
No
9
Switch 2
1. High
2. Low
3. Pulse
4. Rect wave
High
No
10
Threshold
AU
–4.0000 to 4.0000 or
variable, depending on
output selection
Yes
Table 3–3:
Timed event parameters (continued)
Number Event
Units
Range or default
Specify
channel