Making electrical power connections
2-15
•
The maximum range of the detector time constant setting depends on
the wavelength mode and data rate selected. See the
“Primary and secondary function (method) parameters” on page 3-19
•
Empower allows the detector to operate in both single and dual
wavelength modes with a wavelength range of 190 to 700 nm, and up to
4.0 AUFS.
To connect Ethernet cables from the detector to a Waters data system:
1.
Connect the single receptacle end of the Ethernet cable to your data
system by attaching the cable to the network adapter (laboratory
acquisition and control environment, or LAC/E).
Note:
The Ethernet cable is supplied with the Waters data system.
2.
Connect the other end of the cable to the Ethernet connector on the
detector rear panel.
Ethernet connections to the detector in a Waters Empower System
Tip:
When connecting the detector to a data system, all detector parameters
not configurable by the data system in use defer to local control.
Caution:
The maximum total cable length between Ethernet devices
in a system is 20 meters (65 feet). The maximum recommended cable
length between two Ethernet devices is 4 meters (13 feet). Longer
total cable lengths can cause intermittent Ethernet communication
failures.
busLAC/E or LAC/E32 card
Ethernet cable
Ethernet
connector
2489
Detector
workstation
Empower
Summary of Contents for 2489
Page 8: ...viii...
Page 32: ...1 16 Theory and Principles of Operation...
Page 134: ...3 66 Preparing the Detector...
Page 193: ...Warnings that apply to all Waters instruments A 9...
Page 204: ...B 6 Detector Specifications...
Page 220: ...D 14 Solvent Considerations...