Connecting other devices
2-35
Inject-start connections
To program the start of an active method, connect the autosampler’s inject-
start terminals to the detector’s inject-start inputs (see the following table and
figure).
The following figure illustrates the connections between the detector and the
autosampler. Use any available pair of inject-start terminals on the
autosampler.
Inject-start connections for the 717plus Autosampler
Inject-start connections for the 717plus Autosampler
717plus Autosampler terminal
2475 detector (A inputs)
Inject Start + (any one of many paired with
+
)
Pin 1 Inject Start +
Inject Start – (any one of many paired with
–)
Pin 2 Inject Start –
123456
7
8
9
10
11
12
13
14
15
717plus Autosampler
2475 detector
A (inputs and outputs)
1
2
3
4
5
6
7
8
9
10
+
−
+
−
+
−
+
−
Inject start 1
Inject start 1
Ground
Lamp on/off
Lamp on/off
Chart mark
Chart mark
Ground
Auto zero
Auto zero
Red
Black