Making signal connections
2-9
Multiple Waters ethernet instrument connections
Making inject-start signal connections
The Ethernet data system or controller used with the 2475 detector must
receive an inject-start signal from the autosampler or manual injector to
initiate the data collection and time-based programs.
The following table summarizes the inject-start connections for different
system configurations
2475 detector inject-start connections
Inject-start output source
Inject-start input connection
(on the 2475 detector, connector A)
Waters 700 Series Autosampler
Inject Start + / –
Waters Alliance Separations Module Inject Start
Waters manual injector, or
third-party manual injector or
autosampler
Inject Start + / –
TP02075
Switch
eSAT/IN module
Analog input 1
Analog output
2475
detector
ZQ/EMD
1000
100-base-T
ethernet cable
Workstation
Instrument
network card
Analog input 2