Making signal connections
2-5
Making tubing connections
To make tubing connections:
1. Bottom each tubing end in the column outlet, detector inlet, or detector
outlet fitting.
2. Seat each ferrule by tightening the compression screw 1/2-turn past
finger-tight.
Tip:
To ensure accurate verification during installation, be sure to pump fresh,
degassed, and filtered 100% water through the flow cell before powering it.
Making signal connections
See also:
Ethernet Instrument Getting Started Guide
.
The following figure shows the rear panel location of the connectors used to
operate the detector with external devices.
Warning:
To avoid electrical shock, power-off and unplug the detector
before making signal connections.