C-10 Solvent Considerations
UV cutoffs for common solvents
The following table shows the UV cutoff (the wavelength at which the
absorbance of the solvent is equal to 1 AU for some common chromatographic
solvents). Operating at an excitation wavelength near or below the cutoff
increases baseline noise due to the solvent’s ability to absorb excitation light
energy.
UV cutoff wavelengths for common chromatographic solvents
Solvent
UV Cutoff (nm)
Solvent
UV Cutoff (nm)
1-Nitropropane
380
Ethylene glycol
210
2-Butoxyethanol
220
Iso-octane
215
Acetone
330
Isopropanol
205
Acetonitrile
190
Isopropyl
chloride
225
Amyl alcohol
210
Isopropyl ether
220
Amyl chloride
225
Methanol
205
Benzene
280
Methyl acetate
260
Carbon disulfide
380
Methyl ethyl
ketone
330
Carbon
tetrachloride
265
Methyl isobutyl
ketone
334
Chloroform
245
Methylene
chloride
233
Cyclohexane
200
n
-Pentane
190
Cyclopentane
200
n
-Propanol
210
Diethyl amine
275
n
-Propyl chloride 225
Dioxane
215
Nitromethane
380
Ethanol
210
Petroleum ether
210
Ethyl acetate
256
Pyridine
330
Ethyl ether
220
Tetrahydrofuran
230
Ethyl sulfide
290
Toluene
285
Ethylene
dichloride
230
Xylene
290