3-58
Using the Detector
Parameters used for sample and zero-scans
The following table provides the defaults and ranges for all parameters for the
sample and zero-scans.
Sample and zero-scan parameters
Parameter
Screen
Units
Range or
default
Type
1
n/a
Sample scan: 1
Zero-scan: 2
Default: 1
λ
range
2
nm
Range: 200 to
900 nm
Default: 200 or
210 nm
Pace
3
nm/min
Range: 30 to
1000 nm/min
Default: 100
nm/min
EUFS
3
EU
Range: 1 to
100,000
Default: Last
number entered
Tick mark
(Mark each nm)
4
nm
Range: 10 to
100
Default: Last
number entered
Gain
2
n/a
Range: 1 to
1,000
λ
other
2
nm
Default: 200 or
210 nm
Monochromator
scan type
2
n/a
Excitation or
emission