June 11, 2015, 715004754 Rev. A
Page 107
Table 7–3:
Abnormal baseline troubleshooting
Symptom
Possible cause
Corrective action
Baseline drift, rapid
Column not equilibrated
Equilibrate column.
Detector not allowed to
warm up
Allow detector to warm up
until baseline is stable.
Warmup time varies based
on sensitivity.
Solvent contaminated or
not HPLC grade
Use fresh solvent.
Solvent not properly
degassed (rapid or slow
drift)
Degas the solvent. Cap the
solvent reservoir.
Flow cell can possibly
contain an air bubble
Properly purge the
system.
Tubing contaminated
Clean the tubing using the
procedure in
Flow fluctuations (rapid or
slow drift)
Correct pump problems,
replace pump seals,
ensure that valves are
operating properly.
Improper ambient
environment
Verify and correct ambient
temperature and humidity.
Baseline drift, slow
Solvent contaminated
Use fresh, degassed
solvent.
Ambient temperature
fluctuations
Stabilize operating
environment temperature
enough to allow full
equilibration. Keep the
system away from air
conditioning vents,
random breezes, and
direct sunlight.
Flow cell can contain an air
bubble
Properly purge the
system.
Dirty flow cell
Clean flow cell (see
Baseline drift, descending
Leaky flow cell
Call Waters Technical
Service.
Flow cell can contain an air
bubble
Properly purge the
system.
Summary of Contents for 2414
Page 2: ...June 11 2015 715004754 Rev A Page ii...
Page 14: ...June 11 2015 715004754 Rev A Page xiv...
Page 42: ...June 11 2015 715004754 Rev A Page 42...
Page 56: ...June 11 2015 715004754 Rev A Page 56...
Page 62: ...June 11 2015 715004754 Rev A Page 62...
Page 92: ...June 11 2015 715004754 Rev A Page 92...
Page 114: ...June 11 2015 715004754 Rev A Page 114...