7.3.1 Hardware troubleshooting
The following table describes hardware troubleshooting for the detector.
7.3.2 Chromatography troubleshooting
Chromatography troubleshooting helps isolate and correct the causes of these
problems:
•
Abnormal baseline (drift, noise, or cycling) (see the table on
).
•
Erratic or incorrect retention times (see the table on
•
Poor peak resolution (see the table on
).
•
Incorrect qualitative/quantitative results (see the table on
).
Table 7–2:
General system troubleshooting
Symptom
Possible cause
Corrective action
Detector inoperative
Fuse blown
Verify the front panel display is
operational; if it is not, replace
the AC rear panel fuse.
No power at outlet
Confirm that the ac outlet is
energized by connecting
another electrical unit known
to be in working order and
determining whether it
operates.
front panel display fails
to illuminate
Broken electrical
connection
Inspect electrical connections.
Fuse blown
Inspect and, if necessary,
replace fuse(s).
Bad LCD or control board
Contact Waters Technical
Service.
front panel displays odd
characters
Faulty EPROMs
Bad LCD control board
Contact Waters Technical
Service.
IEEE-488 problems
IEEE-488 410 mode
configuration disabled
and/or IEEE address set
incorrectly
Set the configuration page
correctly.
Bad IEEE-488 cable
Check the IEEE-488 cable.
Replace the IEEE-488 cable.
Keypad not functioning Keypad defective
Cycle power to the detector
and run the keypad diagnostic
test.
Contact Waters Technical
Service.
Analog output incorrect RIU-FS setting changed
Reset the RIU-FS setting in
RIU mode.
Summary of Contents for 2414
Page 2: ...June 11 2015 715004754 Rev A Page ii...
Page 14: ...June 11 2015 715004754 Rev A Page xiv...
Page 42: ...June 11 2015 715004754 Rev A Page 42...
Page 56: ...June 11 2015 715004754 Rev A Page 56...
Page 62: ...June 11 2015 715004754 Rev A Page 62...
Page 92: ...June 11 2015 715004754 Rev A Page 92...
Page 114: ...June 11 2015 715004754 Rev A Page 114...