B-7
The following table lists Teflon
®
AF chemical compatibility test results from
Waters.
Tip:
Additional materials in the flowpath are PPS and PTFE.
Output
QSPI serial data
Eluent connections
1/4-28 reversed ferrule fittings;
2 inlet and 2 outlet fittings
Vacuum chambers
One chamber/eluent, two channels
standard
<0.5 mL internal volume/channel
Tubular membrane
1/vacuum chamber, Teflon AF
Vapor exhaust
1/8-inch inside diameter ID
×
1.8 meter
tubing
Vacuum source
Built-in, 2-head diaphragm pump, solvent
resistant
Vacuum sensor
Detects vacuum from 0 to 15.6 psiA
Maximum pressure on inlet
and outlets
70 kPa (0.7 bar, 10 psi)
a. High-density polyethylene
b. polytetrafluoroethylene
c. ethylenetetrafluoroethylene
Teflon AF chemical compatibility test results
Compatible
a
Incompatible
b
Acetonitrile
Fluorinert
®
Chloroform
Fomblin
®
Dimethyl acetamide
Freon
®
Dimethyl formamide
Galden
®
Dimethylsulfoxide
HFIP
Ethanol
Perfluorinated solvents
Ethylacetate
Integral vacuum degasser operational specifications (Continued)
Item
Specification