Identifying chromatographic problems
6-9
Identifying chromatographic problems
The following table is a guide to troubleshooting and correcting
chromatographic problems.
Correcting chromatographic problems
Symptom
Possible cause
Corrective action
Erratic retention
times
Air bubble in pump head
Degas all eluents and
prime pump (see
).
Malfunctioning check
valves
Replace pump check
valves (see
Leaking pump seals
Replace pump seals (see
).
Leaks around fittings
Tighten fittings
Separation chemistry
Check eluent and column.
Clogged eluent filters
Replace filters.
Increased retention
times
Incorrect flow rate
Change flow rate.
Incorrect eluent
composition
Change eluent
composition.
Column heater not turned
on
Turn column heater on.
Incorrect eluent
Use correct eluent.
Column contaminated
Clean or replace column.
Incorrect column
Use correct column.
Air bubble in pump head
Degas all eluents and
prime pump (see
).
Clogged eluent filter
Replace filter.