6-8
Troubleshooting
Identifying and correcting noises
The following table is a guide to troubleshooting and correcting noises in the
Waters 1500-series pump.
Identifying noises
Symptom
Possible cause
Corrective action
Click or loud
snap
Binding plunger seal If click does not stop and cannot be
isolated to one pump head, replace
the plunger seals, one at a time (see
Worn indicator rod
spring
Replace the spring.
Squeak
Plunger seals dry
Wet plunger with appropriate
eluent through pump head access
holes.
Binding plunger seal Replace plunger seal assembly (see
Improper plunger
seal
Install correct pump plunger seal
(see
Binding piston
indicator rod
Replace indicator rod.