5-26
Maintaining the HPLC Pump
Replacing the rear panel fuses
Required materials
•
Screwdriver, flat-blade, small
•
Replacement fuses
To replace a blown or faulty fuse:
•
Power-off the pump, and then remove the power cord from its connector.
•
Use the small flat-blade screwdriver to remove the fuse holder located
just above the power cord connector.
1. Remove and discard the old fuses.
2. Insert the new fuses into the fuse holder.
3. Insert the fuse holder into the receptacle and gently push until it locks
into position.