Connecting the liquid waste line
December 6, 2012, 71500126802IVD Rev. C
179
Connecting the liquid waste line
Required material:
Chemical-resistant, powder-free gloves
To connect the liquid waste line
1. Place a suitable waste container below the instrument.
2. Slide a drain line over the barbed fitting of the drain (located at the
bottom of the instrument).
Warning:
The waste line and connection can be contaminated
with biologically hazardous materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Caution:
To avoid distorting the drip tray or causing the drain
cup to leak, restrain the drain cup when attaching or removing
the waste line.
TP02684
Drain cup
Barbed fitting
Drain line