Quality control
Routinely run three quality-control samples that represent subnormal,
normal, and above-normal levels of a compound. Ensure that quality-control
sample results fall within an acceptable range, and evaluate precision from
day to day and run to run. Data collected when quality-control samples are
out of range might not be valid. Do not report these data until you are certain
that the instrument performs satisfactorily.
When analyzing samples from a complex matrix such as serum/plasma, whole
blood, and so on, note that the matrix components can adversely affect LC/MS
results, enhancing or suppressing ionization. To minimize these matrix
effects, Waters recommends you adopt the following measures:
•
Prior to the instrumental analysis, use appropriate sample
pretreatment such as protein precipitation, liquid/liquid extraction
(LLE), or solid phase extraction (SPE) to remove matrix interferences.
•
Whenever possible, verify method accuracy and precision using
matrix-matched calibrators and quality-control samples.
•
Use one or more internal standard compounds, preferably isotopically
labeled analytes.