General troubleshooting
5-13
Generate Test Peaks diagnostic function
The Generate Test Peaks diagnostic function allows you to override sample
signal input with a simulated Gaussian peak.
To use the generate test peaks function
1. Press DIAG 8, Generate test peaks. If the Generate Test Peaks
diagnostic function is disabled, the Test peaks enabled diagnostic screen
appears.
Test peaks enabled diagnostic screen
2. To disable the Generate Test Peaks function, press Enter.
General troubleshooting
This section suggests possible causes of errors and recommends
troubleshooting actions. Keep in mind that the source of apparent detector
problems can actually be the chromatography, or it can involve other system
instruments.
Most detector problems are relatively easy to correct. If, after running the
diagnostic functions applicable to your problem and troubleshooting the
detector, you cannot correct an error condition, contact Waters’ Technical
Service department.
Power surges
Power surges, line spikes, and transient energy sources can adversely affect
detector operations. Be sure that the electrical supply used for the detector is
properly grounded and free from any of these conditions.