Cleaning the nebulizer ultrasonically
4-9
8. Align the two pins inside the desolvation chamber with the grooves in
the new nebulizer. The quick-disconnect tubing fitting will be at the
twelve o’clock position.
Pins inside the desolvation chamber
9. Push the nebulizer into the nebulization chamber, and turn it clockwise
until it snaps into place.
10. Insert the gas inlet tubing into the quick disconnect-tubing fitting on the
right side of the nebulizer.
11. Reconnect the solvent inlet tubing.
12. Power-on the detector.
See also:
“Selecting the type of nebulizer”
Cleaning the nebulizer ultrasonically
To clean the nebulizer
1. Stop the liquid flow, and remove the solvent inlet line.
Pins