4-6
Maintaining the Detector
8. Replace the left-hand, front panel cover.
9. Power-on the detector, and enter the new lamp information (see
).
Replacing the nebulizer
Required material
Nebulizer
To replace the nebulizer
1. Stop the liquid flow.
2. Power-off the detector, and disconnect the power cable from the rear
panel.
3. Remove the left-hand, front panel cover.
Warning:
To avoid burn injuries, do not touch the nebulizer until
its temperature cools to less than 30 °C, as displayed on the
detector home screen. If its temperature exceeds 30 °C, let the
nebulizer cool in one of two ways before touching it:
• Wait 30 minutes after powering-off the detector.
• Wait 10 minutes after specifying Cool in the Nebulizer and
Drift Tube Temperature Control Home screen (see
).