9.
INPUT ADVANCED SETTING
Here, the items of input setting which are not explained in the “BASIC SETTING” are explained.
StrainInputA, StrainInputB, ProcessInputA, ProcessInputB
AnalogFilter(Only StrainInputA and StrainInputB)
Select a low pass filter for removing high frequency noise of input signal.
Please set according to your environment.
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB]
⇒
[AnalogFilter]
Setting value: OFF,30Hz,300Hz(Default),600Hz
*If this item is changed, measurement will be invalid for about 50 ms after back to measurement display.
SamplingRate
Select a sampling rate. Max rate of 1ch input products is 4000 samlping per sec(sps), and of 2ch input products is
2000 sps.
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB][ProcessInputA][ProcessInputB]
⇒
[SamplingRate]
Setting value:4000sps (Default for 1ch input products), 2000sps (Default for 2ch input products), 1000sps,
500sps, 200sps, 100sps, 50sps, 20sps, 10sps, 5sps, 2sps Every1s, Every2s, Every5s, Every10s
MoveAve
Select the number of moving average.
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB][ProcessInputA][ProcessInputB]
⇒
[MoveAve]
Setting value:
None, 2times, 4times, 8times, 16times, 32times(Default), 64times
InputCorrect
Select linearize ON or OFF.
Linearize is a function to correct by maximum 16 points of polygonal line approximation when the sensor
characteristics are nonlinear.
Operation:[MENU]
⇒
[Input]
⇒
[StrainInputA][StrainInputB][ProcessInputA][ProcessInputB]
⇒
[InputCorrect]
Setting value:
None(Default), Linearize
* Calculation processing of linearize is performed after calculation processing of “DispLoCut”.
(“DispLoCut” is to be described later)