![WATANABE WPMZ-1 Instruction Manual Download Page 63](http://html1.mh-extra.com/html/watanabe/wpmz-1/wpmz-1_instruction-manual_3499830063.webp)
Graphical Digital Panel Meter WPMZ-1 INSTRUCTION MANUAL
62
IM-0882-03
15.
DIAGNOSIS
This section explains the items relate to diag setting of diag/log setting.
(The items related to log setting are in other section.)
・
InputDiag:
You can check whether the output of the sensor is performing normally or check ON / OFF of the
external controls.
・
OutputTest:
You can try comparisonjudgemant outputs, pulse outputs and other option outputs for operation
confirmation.
InputDiag
DC InputA/B, ProcessInputA/B
Display process input level in percentage and actual value.
Operation:[MENU]
⇒
[Diag/Log]
⇒
[InputDiag]
⇒
[DC InputA/B] [ProcessInputA/B]
Display:
InputRatingPercent: (Input level in percentage.)
InputActualValue: (Input level in actual value.)
ExternalCtrl
Display statuses of each external controls.
Operation:[MENU]
⇒
[Diag/Log]
⇒
[InputDiag]
⇒
[ExternalCtrl]
Setting:
Terminal 1 ON / OFF
Terminal 2 ON / OFF
Terminal 3 ON / OFF
Terminal 4 ON / OFF
Terminal 5 ON / OFF
*Statuses are reflected in real time.
OutputTest
CompareAL1~AL4
You can arbitrarily turn on / off the comparison judgment outputs in order to check connected device.
Operation:[MENU]
⇒
[Diag/Log]
⇒
[OutputTest]
⇒
[CompareAL1
~
AL4]
Setting:
CompareAL1 TestOutput:
Disable (Initial state) / Enable
TerminalOutput:
ON Output/ OFF Output (Initial state)
CompareAL2
~
4
Same as above.
*The output enables as soon as TestOutput turn on.
TestOutput is enabled until it is disabled or power off.
TestOutput automatically turn off when re-power on.
GoOutputA/B
You can arbitrarily turn on / off the GO outputs in order to check connected device.
Operation:[MENU]
⇒
[Diag/Log]
⇒
[OutputTest]
⇒
[GoOutputA/B]
Setting:
GoOutputA TestOutput:
Disable (Initial state) / Enable
TerminalOutput:
ON / OFF (Initial state)
GoOutputB Same as above.
*The output enables as soon as TestOutput turn on.
TestOutput is enabled until it is disabled or power off.
TestOutput automatically turn off when re-power on.
AnalogOutput
You can arbitrarily set analog output by 10% steps in order to check connected device.
*This function is for only products that have analog output option.