(1)The application of voltage or current exceeding its maximum 

allowable value to the input terminals may result in instrument 
damage. 

(2)The supply of power out of its allowable range may cause fire, 

electric shock or instrument failure. 

(3)The content of this manual may subject to change without prior 

notice for product improvement. 

(4)This manual is carefully prepared. However, if any question  

arises, or any mistake, omission or suggestion is found in the 
content of this manual, contact your nearest our sales agent. 

  (5)Make this manual available easily anytime. 
 
 
 

 
1.Specifications 

  ●DC Voltage Measurement 

Model & 

Range Code

 

Measuring  

Range 

Maximum 

Resolution 

Input 

Impedance 

Maximum 

Allowable 

Input voltage 

AP-101-11 

±199.9mV 

100μV 

100MΩ 

±250V 

AP-101-12 

±1.999V 

1mV 

100MΩ 

±250V 

AP-101-13 

±19.99V 

10mV 

10MΩ 

±250V 

AP-101-14 

±199.9V 

100mV 

10MΩ 

±500V 

AP-101-15 

±700V 

1V 

10MΩ 

±700V 

Accuracy: ±(0.1% of rdg +1digit)(at 23℃±5℃ 35 to 85%  RH) 
         [±(0.3% of rdg +1digit)only for AP-101-15] 
  ●DC Current Measurement 

Model & 

Range Code

 

Measuring  

Range 

Maximum 

Resolution 

Internal 

Resistance 

Maximum 

Allowable 

Current 

AP-101-21 

±199.9μA 

100nA 

1kΩ 

±10mA 

AP-101-22 

±1.999mA 

1μA 

100Ω 

±50mA 

AP-101-23 

±19.99mA 

10μA 

10Ω 

±150mA 

AP-101-24 

±199.9mA 

100μA 

1Ω 

±500mA 

AP-101-25 

±1.999A 

1mA 

0.1Ω 

±3A 

Accuracy: ±(0.2% of rdg +1digit)(at 23℃±5℃ 35 to 85%  RH) 
         [±(0.3% of rdg +1digit)only for AP-101-25] 

 

 

●Model Configuration 
 (Example) 
 
 
 
 
 
 
 
 
 

 
 
 

The above standards do not apply to the panel meter 
with “15 range”. 

2.Common Specifications 

Measuring function 

: DC voltage measurement. 

Specify one model from among 9 
models. 

Operation method 

:Dual slope integration 

Input circuit 

:Single-ended type 

Input bias current 

:50pA(Typical) 

Sampling speed 

:2.5 times/sec. 

Noise elimination 

:More than 40dB(50/60Hz)(Typical) 

Max. No. of display 
digits 

:1999 

Overrange alarm 

:1999 or -1999 flashes when an input 

exceeding the maximum display range 
is applied. 

Display 

:LED(Light Emitting Diode)numeric 

element 

  Height; 14.2mm red 

Polarity 

:Automatic Polarity selection 

Polarity display 

:“-“ is displayed automatically if 

input signal becomes negative. 

External control 

:・External display hold 

A negative signal of 0V or a contact 
signal. 

・External start 

A positive pulse from 0V to +5V for 
more  than  400ms.  or  a  contact 
signal(open) 

Decimal point 

:Settable to any digit position 

Operating 
temperature/humidity 

:0 to 50℃/35 to 85%RH 

(Nodew-Condense) 

Storage temperature 

: -10 to 70℃,60%RH max. 

Power supply 

:For AC, 90V to 132V 50/60Hz 
          Approx.2.0VA(at 100V) 
          180V to 264V 
  For DC, 5V DC ±5% 150mA(MAX) 
          Isolated 
          24V DC±20% 40mA(MAX) 

Isolated 

Dimensions 

:  96mm(W)×48mm(H)×73mm(D) 

Weight 

: For AC; Approx.150g(Mainframe) 

    For DC; Approx.85g(Mainframe) 

Dielectric strength 

: For AC; 2100VAC for 1 minute between 

power supply terminal and input 
terminal/COMMON 

  For DC; DC±500V between input (LO) 

terminal and power terminal(0V) 

Insulation 
resistance 

:More than 100MΩ  at 500VDC 
  between power terminal and input 

terminal 

Accessories 

:Instruction manual, connector 

Standards 

: EN61326-1:2006 

EMI:Class A   
EMS:Industrial locations 
※Cable length:30m or less 
※In the case of DC drive DC power 
supply port: DC connections between 
parts of equipment 
EN61010:2001 
The above standards do not apply to 
the panel meter with “15 range”. 

Setup enviroment 

: Installation categoryⅡ, 

Pollution degree2 

Altitude 

: 2000m max. 

 

Caution 

Attention 

The  mark  on  the  label  indicates  the  measuring   

range  in  the  specification  of  item  1. 

Do not disassemble or touch the interior  

while the power is ON. 
This may cause an electric shock. 

INSTRUCTION MANUAL 
DIGITAL PANEL METER 
MODEL AP-101 Series 

Warning 

Incorrect  handling  may  cause 

death or injury. 

(3/4) 

On the AP-101, connect the control inputs(HOLD,Decimalpoint)to 
base-isolated devies when the input signal is 70V DC or more. 

WATANABE ELECTRIC INDUSTRY CO., LTD.

11.±199.9mV

12.±1.999V

13.±19.99V

14.±199.9V

Series

Basic Model

Range code

Power supply

4. DC24V

3. DC 5V

2. AC180V to 264V

1. AC 90V to 132V

21.±199.9μA

22.±1.999mA

23.±19.99mA

24.±199.9mA

25.±1.999A

AP-101-11-1

15.± 700V