GEN.S.19/167e 

DECEMBER 2019 

MR5501 

MR Conditional Adjustable Height 

Patient Trolley 

OPERATORS MANUAL 

Summary of Contents for MR5501

Page 1: ...GEN S 19 167e DECEMBER 2019 MR5501 MR Conditional Adjustable Height Patient Trolley OPERATORS MANUAL...

Page 2: ...ormation regarding operation service and maintenance If you have any problems in the meantime or would like any advice about this or any other MR products from the Wardray Premise range please contact...

Page 3: ...ed conditions At a minimum address the conditions of the static magnetic field the switched gradient magnetic field and the radiofrequency fields Additional conditions including specific configuration...

Page 4: ...A service contract package is available and details are available upon request from our Service Department on 44 0 20 8398 9911 For international customers please contact your local Wardray Premise re...

Page 5: ...tional Directional Locking Castor see page 9 Backrest gas cylinder Serial Number under backrest Brake Lever Plunger IV Pole stored on left of head Push bar Thumb screw for IV Pole Pull Handle for Tren...

Page 6: ...raise until it stops repeat until desired height is reached or top height is achieved Fig 2 To lower place foot on black rubber and push down until you feel the stop then push further and table top w...

Page 7: ...the horizontal position the upward movement is power assisted Backrest The backrest is operated by squeezing the handle located under the right hand side of the backrest board at the head end as shown...

Page 8: ...it stops To raise the cot sides hold the top rail in the centre and lift until the plunger engages To operate Flag Plunger To lower the cot sides disengage the black flag plunger by pulling and turni...

Page 9: ...ocking Castor this is denoted by the green locking lever To engage a directional locking castor 1 Stand at the head end of the trolley push the trolley forward by approximately 1 metre This will ensur...

Page 10: ...Holder An Oxygen Cylinder holder is also optional suitable for a size E cylinder it is usually fitted on the right hand side of the base Fig 1 Fig 1 WARNING THE OPERATOR MUST ENSURE THE OXYGEN CYLIND...

Page 11: ...llows the pole both to be locked vertically for use and horizontally for storage as shown Fig 1 and Fig 2 Fig 2 Fig 1 The IV Pole is manufactured from 316 stainless steel and is stamped with the seria...

Page 12: ...p Push the trolley by the cot sides these are for patient safety only Place a weight of over the equivalent of 34 Stone 220Kg 485lbs onto the trolley Spill oil or other liquids chemicals on the mattre...

Page 13: ...a couple of times before using check for any leaks especially if the trolley has not been used for a long period of time Check the cot sides for functionality and damage especially the cot side bracke...

Page 14: ...If required the product can be washed down with hand hot sodium hypochlorite solution up to 1 or 10000 parts per million chlorine combined with a detergent Alternatively follow your local guidelines o...

Page 15: ...15...