Technical data
STEAM GENERATOR
Pressure safety valve
72.51 psig (5 bar)
Safety thermostats
446 °F (230 °C)
DISTILLED OR DEMINERALIZED WATER
Water quality
See the Test instructions service book (conductivity < 10μS/cm)
Average water consumption
0.17 to 0.2 gal/cycle (0.65 to 0.75 litres/cycle)
Tank volume
Clean water 1.7 gal (6.4 l)
Min. water charge (clean water): 0.29 gal (1.1 l)
Used water 1.77 gal (6.7 l)
Recommendations for validation
TEST VALIDATION POINTS
Part
Description
Hottest points
Coldest points