8
MN-111 | Instructions for use | MN-111 W&H Med ENG Rev11 | 2019/01/25 | © 2019 W&H Sterilization Srl
Introduction
All rights reserved in all countries.
All drawings, images and texts contained in this manual are the
property of the manufacturer. Even partial duplication of drawings,
images or text is prohibited.
The information contained in this document is subject to change
without prior notice.
Use restrictions
INTENDED USE
The MN-111 is designed for pressurized steam sterilization of
medical and pouched dental instruments, including dental
handpieces, porous and hollow loads.
Key program features, including sterilization time, temperature and
recommended load type are listed in the following table:
Program
Type of Load and
Load weight
Sterilization
Temperature
Sterilization
Time
Drying Time
(recommended)
Pouches
Medium
Load
Instruments and
dental handpieces,
up to 3 lbs (1.4 kg).
270 °F (132
°C)
4 minutes
25 minutes
Pouches
Large Load
Instruments and
dental handpieces,
up to 6 lbs (2.7 kg).
270 °F (132
°C)
4 minutes
30 minutes
Program
Type of Load and
Load weight
Sterilization
Temperature
Sterilization
Time
Drying Time
(recommended)
Wrapped
Cassettes
Instruments and
dental handpieces,
up to 14 lbs (6.4
kg).
270 °F (132
°C)
4 minutes
40 minutes
Unwrapped
Instruments and
dental handpieces,
up to 18 lbs (8.2
kg).
270 °F (132
°C)
4 minutes
8 minutes
Low
Temperature
Textiles, up to 4.4
lbs (2 kg), or
instruments and
dental handpieces
requiring low
temperature, up to
5 lbs (2.3 kg).
250 °F (121
°C)
30 minutes
30 minutes
See "Sterilization programs" on page 85 for the full list of key
program features, including sterilization time, temperature, drying
time and recommended load type.